Horizontal Image Area
2849636
2849636
ENRAF-NONIUS ENDOLASER 120
Laser therapy unit
Gallery
2849637
2849637
ENRAF-NONIUS ENDOLASER 120
Laser therapy unit
HIGH PERFORMANCE THERAPEUTIC LASER
The 905 nm GaAs laser probe delivers pulses in nanoseconds, producing average powers of 100 mW and peak powers up to 100 W per diode. This results in a higher concentration of light energy (I₀), or photon density at tissue depth, without the risk of burning tissue.
PROVEN EFFECTIVENESS
The clinical application of light – produced by laser diodes with a power output up to 1600 mW - is scientifically well documented. The 905 nm (GaAs) pulsed laser is the most versatile for deep lying structures (joints, vertebrae, back (muscles) whereas the 808 nm (GaAlAs) is more appropriate for wound healing and tissue repair.
Scientific evidence continues to demonstrate that pulsed light does have biological and clinical effects that are different from those of continuous wave (CW) light. Several studies revealed that the LLLT in pulsed wave mode of operation can better penetrate through the melanin and other skin barriers, supporting the hypotheses that pulsing is beneficial in reaching deep target tissue and organs.
EVIDENCE BASED CLINICAL GUIDELINES
The Endolaser provides the user with dosage recommendations for the effective treatment of pain and various musculoskeletal disorders.
MULTIPLE LASER PROBES AVAILABLE
For the treatment of smaller surfaces probes, there is a choice of single laser probe with a total power up of 100mW and 500mW are available. For the treatment of larger areas, the cluster probe, combining 4 diodes with a total up to 1600 mW, can be used. The Endolaser automatically recognizes the type of probe that is connected.
QUICK ACCESS POWER BUTTONS
The output power of the Endolaser 120 can be easily set at different levels for easy dosage settings.
TARGET LIGHT
When treating a patient without probe-skin contact, the red target light helps the therapist to focus on the area to treat. The target light contributes to a safe treatment by indicating the irradiation area of the laser beam.
ERGONOMIC PROBES
The ergonomic probes can be used with or without skin contact. Each probe is equipped with a start-pause button for easy control. The indication light on top of the probe indicate the emission state
(laser running or laser-ready).
TECHNICAL SPECIFICATIONS
Mains voltage: 100 - 240 Volt
Frequency: 50/60 Hz
Max. power output: 20 VA
Dimensions device (length x width x height): 22 x 16 x 14 cm
Weight unit: 1.6 kilograms
Type of laser diode: GaAs (pulsed) and GaAlAs (continuous)
Preprogrammed clinical protocols: 22
Number of favorites to save: 20
Safety and performance standards
Medical device classification IIa; Rule 9 Annex IX of 93/42/ EEC
This equipment complies with all requirements of the Medical Device Directive (93/42/EEC).
IEC 60601-1 General requirements for the safety of electrical medical systems.
Safety class according to IEC 60601-1 Electrical safety class II
Applied parts Type B Applied part.
Classification of laser according to IEC60825-1 3B
STANDARD ACCESSORIES
2x Laser Safety Eyewear (goggles)
Infosheet &User manual (on CD)
Mains cable 230V-EUR
1x Probe holder (premounted)
1x Doorswitch cable
OPTIONAL ACCESSORIES OUTPUT PEAKPOWER
LP100P Laser probe 100 mW, super pulsed laser diode (905 nm) 100 mW 100 W
LP500C Laser probe 500 mW, continuous laser diode (808 nm) 500 mW
Cluster probe 4x100 mW, continuous laser diode (808 nm) 4 x 100 mW
Cluster probe 4x400 mW, continuous laser diode (808 nm) 4 x 400 mW
Laser Safety Eyewear (set of goggles).